S021: Test Step 4 - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016

S021: Test step 4

8.1.17.2
Initiate the packing position for the sensor unit
1. Call service routine S021.4 [ → 249].
An inactive progress indicator in selection field 1 and the
message FFFF in selection field 2 signal that the system is ready
to move the sensor unit.
2. Touch the T key (A).
The sensor unit is moved to the packing position.
During this time, a progress indicator is displayed.
When the packing position is reached, "0000" is displayed in
selection field 2.
3. Exit the service routine [ → 252].
8.1 Perform service routines via the control panel
8 Service routines
303