Sirona ORTHOPHOS SL 2D Service Manual page 196

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
196
7.2.1.3.2
Manual adjustment: Pan - Symmetry
The manual adjustment procedure is similar to the one for automatic
adjustment [ → 182]. The only difference is that the default adjustment
values calculated automatically in the
overwritten by adjustment values calculated manually.
1. Start the
described in the chapter entitled "Automatic adjustment: Pan -
Symmetry [ → 193]".
2. After taking the exposure, measure the distances S1, S2 and S3 with
the toolbar measuring ruler [ → 166].
If possible, measure in the lower area of the needles since they may
be slightly bent after several uses.
Tip: For better measurement, you can also edit [ → 166] the exposure
using the toolbar functions.
3. Overwrite the adjustment values for S1, S2 and S3 in the text boxes
"Symmetrie adjustment"
of the
For information about the shifting direction (input of +/- sign in the
menu), see the chapter entitled "Shifting direction of the exposed
image area [ → 174]".
NOTICE! Use points as decimal separators!
4. Continue with the adjustment procedure as described in the chapter
entitled "Automatic adjustment: Pan - Symmetry [ → 193]".
"Symmetrie adjustment"
"Symmetrie adjustment"
menu with the measured values.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
menu are
adjustment procedure as
64 95 258 D3632
D3632.076.01.03.02 09.2016

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos sl 3dOrthophos sl 3d ceph

Table of Contents