Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
x
Description
4
Malfunction or
mechanical obstruction
of unit movements
5
Malfunction during the
exposure or during
exposure preparation.
6
Error during system
self-test.
7
Unrecoverable system
error.
64 95 258 D3632
D3632.076.01.03.02 09.2016
Error group
This error group includes all errors
that indicate problems with the
motor-controlled movements on
the outside of the unit.
This error group includes all errors
resulting from a certain system
action triggered by the user which
could not be performed because a
required (internal) partial function
(software or hardware) is not
ready or fails.
This error group includes all errors
which may occur spontaneously
and without any related operator
action. They may be caused by
system self-tests.
This error group includes all errors
which may occur spontaneously
and without any related operator
action. They may be caused by
system self-tests. In this case it is
absolutely certain that continued
system operation is not possible.
5.3.3

yy - Location

Identifier yy defines the location or logical function unit where the error
has occurred.
The location may be a DX module number standing for an entire HW
function unit, or a logical SW function unit on board DX11 (central
control).
5.3.4

General handling of error messages

Error messages must generally be acknowledged with the R key on the
touchscreen.
If trouble-free operation is possible after the error is acknowledged, then
no further action is necessary.
If error messages occur again or frequently, or if fault-free operation is not
possible, run an error diagnosis (see chapter "Troubleshooting [ → 139]").
In some cases, it may make sense to obtain more information on the
history and frequency of errors via the
section entitled "Opening / saving "Unit information" [ → 59]").
Actions required
● Acknowledge the error message and
make sure that the movements of the unit
are not obstructed.
● Repeat the last procedure step or
exposure.
If the error occurs again ...
● Run an error diagnosis [ → 84].
● Acknowledge the error message.
● Repeat the last procedure step or
exposure.
If the error occurs again ...
● Run an error diagnosis [ → 84].
● Acknowledge the error message.
● Run an error diagnosis [ → 84].
Further operation of the unit is possible.
● Run an error diagnosis [ → 84].
"Device details"
5 Messages
5.3 Error messages
function (see the
83