7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
208
7.2.1.6.2
Manual adjustment: Ceph secondary diaphragm
The manual adjustment procedure is similar to the one for automatic
adjustment [ → 205]. The only difference is that the default adjustment
values calculated automatically in the
diaphragm"
manually.
1. Start the
procedure as described in the chapter entitled "Automatic
adjustment: Ceph secondary diaphragm (standard) [ → 205]"
"Adjustment of Ceph secondary
menu are overwritten by adjustment values calculated
"Adjustment of Ceph secondary diaphragm"
S1
S2
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
adjustment
S3
64 95 258 D3632
D3632.076.01.03.02 09.2016