Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
If pulse/pause monitoring has already been disabled
prematurely, this service routine simply displays the status and
the arrow keys are not shown. The status cannot be changed.
2. To disable pulse/pause monitoring, use the arrow keys (A) to select
the code "00".
The Memory key (B) lights up.
3. Save the setting [ → 251].
PulsePause monitoring is irreversibly disabled.
4. Exit the service routine [ → 252].
8.1 Perform service routines via the control panel
8 Service routines
263