S034: Test step 5

8.1.20.2
Center position of ceph scan secondary diaphragm axis, display/
calibrate
1. If not already done, detach the cover to the cephalometer [ → 37].
2. Call service routine S034.5 [ → 249].
After selecting the service routine, the current offset value for the
center position of the ceph scan
displayed in selection field 1.
3. Touch the T key (A).
The test mode is activated.
In the test mode, travel to the center position of the ceph scan
secondary diaphragm axis is initially performed without offsets.
The T key lights up and indicates the test mode.
4. Move the secondary diaphragm to its front position by touching the R
key (B). The secondary diaphragm can always be moved back and
forth between its front and center positions with the R key.
5. Move the secondary diaphragm to its center position by touching the
R key.
6. Measure distances L2 and L1 with a slide gage.
8.1 Perform service routines via the control panel
secondary diaphragm axis
8 Service routines
is
313