9 Repair
9.12 Sensor unit
O
O
P
388
9. Hold the DCS sensor in place on the frame (K) and carefully remove
the heatsink (L).
NOTE: This can stick slightly to the DX83 board due to the two
thermal conducting foils.
10. Remove the two thermal conducting foils (M).
These are not reused.
M
M
11. Loosen the three screws (N) and remove the DX83 board.
12. Install the new board DX83 with the three screws (N) on the DCS
sensor.
13. Attach the two new HF shields (O) (included in the scope of supply
for board DX83) on the cover plate (P) of board DX83.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
DX83
N
N
N
64 95 258 D3632
D3632.076.01.03.02 09.2016