Sirona ORTHOPHOS SL 2D Service Manual page 259

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
The radiation time corresponds to the selected radiation time.
Important: If you let go of the release button before this radiation
time expires, radiation will be terminated prematurely. The actual
radiation time is not displayed.
When you release radiation during the cool-down interval, a
countdown of the remaining waiting time is displayed in the title
bar of the control panel (automatic exposure blocking).
4. Exit the service routine [ → 252].
8.1 Perform service routines via the control panel
8 Service routines
259

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos sl 3dOrthophos sl 3d ceph

Table of Contents