5 Messages
5.1 Help messages
Help code
Description
"R button, move into starting position"
H3 01
"Change bite block"
H3 07
"R button, confirm exposure data"
H3 20
"Close the door"
H3 21
"Select quadrant"
H3 22
"Select region of exposure"
H3 25
"Plug sensor into Ceph slot"
H4 02
80
5
Messages
The different message texts are displayed ...
● on the touchscreen of the Easypad
● On the display of the remote control
There are 3 groups of message texts:
Help messages (Hx xx):
● Auxiliary messages request an action,
e.g. H3 20 (acknowledge exposure data).
Error messages (Ex yyxx):
● Error messages indicate unit faults.
● The user must take action to eliminate the fault(s).
5.1
Help messages
Help messages are displayed as help codes (Hxxx) on the Easypad
touchscreen or on the remote control display (if available). The codes tell
you how to operate the system if radiation release is not possible.
The following list provides you with an overview of all help codes, their
meaning and the action required to eliminate the corresponding
problems.
IMPORTANT: The measures listed only clear help messages that result
from operator errors. If it is not possible to clear a message by taking the
measures listed, another type of error is the cause. In such cases, you
should run an error diagnosis [ → 139].
Service Manual ORTHOPHOS SL
Actions required
● Press the Return key.
● Unit moves to starting position.
Remove the occlusal bite block.
● Press the Return key.
● Exposure data are confirmed.
● Close door or check door contact.
Then press R key to acknowledge the
message.
● Select quadrant
● Select the center of rotation.
● Check the ceph sensor or plug it into the
cephalometer.
If this message does not disappear even after
the ceph sensor has been plugged in, this
indicates a system error. Perform error
diagnosis as described in the section titled
Troubleshooting [ → 139] .