S018: Test Step 7 - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016

S018: Test step 7

8.1.15.6
NOTE: In order to execute this service routine, the occlusal bite block
First execute SR S017.18
function must be activated via service routine S017.18[ → 288].
Setting the nominal angle for the occlusal bite block (as required by
the customer)
Symbol on the control
panel
Patient symbol key 1 (G) visible
Selection field
1
2
3
1. Call service routine S018.7.
The selected program group is displayed in selection field 1 (to
which the nominal value displayed in selection field 2 applies).
The nominal value (for the program group selected in selection
field 1) is displayed in selection field 2.
The currently measured angle of the bite block plate is displayed
and permanently updated in selection field 3
(possible deflection: -17° to +22°).
The display of patient symbol key 1 (G) indicates whether or not
the bite block is connected (see table).
8.1 Perform service routines via the control panel
Status
Meaning
Occlusal bite block
connected
not visible
Occlusal bite block not
connected
Parameter/Display
Program groups
Nominal angle of the bite block
plate
Measured deflection of the
bite block plate
8 Service routines
Range of values
P1 - P12
BW1 - BW2
VOL1 - VOL3
-13° to +18°
-17° to +18°
293