8 Service routines
8.1 Perform service routines via the control panel
SR*
S021
S021.3
S021.4
302
8.1.17
S021: Service for the packing position
SA**
Function
Service for packing position
No
Initiate the packing position for the diaphragm
No
Initiate the packing position for the sensor unit
* SR=service routine, ** SA=security access
8.1.17.1
S021: Test step 3
Initiate the packing position for the diaphragm
1. Call service routine S021.3 [ → 249].
2. Touch the T key (A).
3. Exit the service routine [ → 252].
An inactive progress indicator in selection field 1 and the
message FFFF in selection field 2 signal that the system is ready
to move the diaphragm.
The diaphragm is moved to the packing position.
During this time, a progress indicator is displayed.
When the packing position is reached, "0000" is displayed in
selection field 2.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016