5 Messages
5.4 List of error messages
Error code
Description
E7 61 12
Unit is not ready for operation
Error code
Description
E6 61 13
Error writing on the persistent
memory of the diaphragm unit.
Error code
Description
E7 61 14
Communication error of the motor
driver on DX61.
Error code
Description
E5 61 15
Drive overtemperature.
Can occur after frequent diaphragm
travel, e.g. during demo operation at
a fair.
Error code
Description
E7 61 17
The installed diaphragm mechanics
do not match the system version.
Error code
Description
E5 61 20
Horizontal motor malfunction during
operation.
Can occur after frequent diaphragm
travel, e.g. during demo operation at
a fair.
126
Actions required
E7 61 12
This error is a sequential fault.
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure and observe causal
error messages.
● Check the CAN bus.
● Replace diaphragm unit.
Actions required
E6 61 13
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
● Replace diaphragm unit.
Actions required
E7 61 14
● Replace diaphragm unit.
Actions required
E5 61 15
● Allow unit to cool down.
If the error occurs repeatedly ...
● Check board DX61; replace the
diaphragm unit, if necessary.
Actions required
E7 61 17
● Replacing the diaphragm unit
Actions required
E5 61 20
● Allow unit to cool down.
If the error occurs repeatedly ...
● Check board DX61; replace the
diaphragm unit, if necessary.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
see
S. [ → 141],
S. [ → 370]
see
S. [ → 370]
see
S. [ → 370]
see
S. [ → 146],
S. [ → 370]
see
S. [ → 370]
see
S. [ → 146],
S. [ → 370]
64 95 258 D3632