8 Service routines
8.1 Perform service routines via the control panel
322

S037: Test step 2

8.1.21.2
Setting the default IP address, default gateway address and default
subnet mask
IMPORTANT
The network address can only be restored to the factory setting (default
value) in fixed address boot mode (STATIC or no DHCP).
Symbol on the control
panel
Patient symbol key 1 (A)
Patient symbol key 2 (B)
Patient symbol key 3 (C)
1. Call service routine S037.2.
Once the service routine has been selected, the network data is
displayed as in test step 1 [ → 320].
The Memory key and the R key also become visible.
The Memory key lights up.
2. Use the patient symbol keys (A, B, or C) to check the network data
existing in the system prior to the restore operation (see table).
The patient symbol key selected in each case lights up.
3. To reset the network data, press the Memory key (E) (R key lights up)
followed by the R key (F).
The default network data (factory default setting) is displayed. To
toggle the display of the various items of network data, proceed
as in test step 1 [ → 320].
4. Exit the service routine.
5. Perform a restart of the unit.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Status
Meaning
lit
The IP address is displayed in
selection field 1*
lit
The default gateway is displayed
in selection field 1
lit
The subnet mask is displayed in
selection field 1
D3632.076.01.03.02 09.2016