8 Service routines
8.1 Perform service routines via the control panel
316

S034: Test step 6

8.1.20.3
Moving to the Ceph packing position
1. Call service routine S034.6 [ → 249].
After calling the service routine, an inactive progress indicator is
displayed in selection field 1. Selection field 2 shows the
characters "FFFF".
2. Touch the T key (A).
The cephalometer moves to the packing position. The procedure
is visualized by an active progress indicator in selection field 1.
When the cephalometer has reached its packing position, "0000"
is displayed in selection field 2.
3. Exit the service routine [ → 252].
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016