9 Repair
9.16 Cable
404
9.16.2

Replacing cables

CAUTION
Switch the unit off before you start replacing cables or removing
connectors.
NOTICE
Be careful not to twist the cables or kink the fiber-optic light guides when
installing them.
Always check the cables before replacing them [ → 153].
The cables are labeled with small flags. They specify the designation and
part number of the cable. The plugs and sockets on the cables are
designated both on the boards and cables. Check the designation when
you pull off the cables.
Some cables feature markings of green adhesive tape. Mark the
corresponding positions on the unit before removing an old cable. Lay the
new cable so that the cable markings again come to rest at the
corresponding positions marked on the unit while removing the old cable.
An overview of all cables can be found in the chapter "Overview of cabling
[ → 26]".
Sirona Dental Systems GmbH