S009: Flash File System; S009: Test Step 4; S009: Test Step 5 - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

8 Service routines
8.1 Perform service routines via the control panel
SR*
S009
S009.4
S009.5
268
8.1.9

S009: Flash file system

IMPORTANT
When the flash file system is formatted, the content of the error logging
memory is lost.
IMPORTANT
After running service routine SR 009.4 a software update to the currently
installed software version must be performed to install the DX11 FGBA
file again.
SA**
Function
Flash file system
yes
Initializing the flash file system
No
Test flash file system
* SR=service routine, ** SA=security access

S009: Test step 4

8.1.9.1
Formatting flash file system
1. Call service routine S009.4 [ → 249].
2. To initialize the flash file system, press the Memory key (A) (R key
lights up), followed by the R key (B).
3. Exit the service routine [ → 252].
8.1.9.2

S009: Test step 5

Test flash file system
Flash file system formatting in progress. This process takes
approx. 5-6 mins and is visualized by a progress indicator. The
end of this process is indicated by the message "0000" in
selection field 2.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016