8 Service routines
8.1 Perform service routines via the control panel
SR*
S032
S032.10
S032.50
304
8.1.18
S032: Sensor test
SA**
Function
Sensor test
No
Self-test of the DCS image data path
No
Self-test of the FP image data path
8.1.18.1
S032: Test step 10
Self-test of the DCS image data path
1. Call service routine S032.10 [ → 249].
2. Press the T key.
3. Exit the service routine [ → 252].
Once the service routine has been selected, "DCS test" is
displayed in selection field 1.
The DCS image data path is tested.
If the image data path is OK, "Success" is displayed in the
selection field 1.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016