8.1.15

S018: Service for height adjustment

SA**
Function
Service for height adjustment
No
Set the maximum travel height
No
Undo the maximum travel height setting
No
Check the height adjustment sensor system
No
Setting the minimum travel height
No
Undoing the minimum travel height setting
No
Nominal angle setting for the occlusal bite block (as required by the
customer)
No
Functional test of the occlusal bite block
No
Calibrating the occlusal bite block
No
Setting the PWM values for the height adjustment creep speed (PWM =
Pulse Width Modulation)
* SR=service routine, ** SA=security access
8.1.15.1

S018: Test step 2

NOTE: Limiting the max. travel height is only possible from a position
value of >1500.
Set the maximum travel height
1. Move the unit to the required maximum travel height by pressing the
Up/Down keys in the user mode on the control panel.
2. Call service routine S018.2.
3. To save the maximum travel height, touch the Memory key (B) (R key
(C) lights up) followed by the R key (C).
8.1 Perform service routines via the control panel
Once the service routine has been selected, the current height
position is displayed in selection field 1.
The Memory key (B) lights up.