8 Service routines
8.1 Perform service routines via the control panel
*
290
4. *For greater clarity the unit is shown in the adjacent image without a
cover.
D
Set the mechanical limit stop on the unit:
Loosen the nut (D) and move mechanical limit stop (E) for the limit
switch until it engages. Tighten nut (D) again.
The next time the UP key is pressed, the unit stops 10 mm below the
limit switch.
5. Exit the service routine.
E

S018: Test step 3

8.1.15.2
Undo the maximum travel height setting
1. Call service routine S018.3.
Once the service routine has been selected, the current height
position is displayed in selection field 1.
The Memory key (B) lights up.
2. To increase the limit for the maximum travel height, first touch the
Memory key (B) (R key (C) lights up) followed by the R key (C).
3. Exit the service routine.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632