Service Manual ORTHOPHOS SL
"Sensor adjustment"
(Pan - adjustment)
64 95 258 D3632
D3632.076.01.03.02 09.2016
7.1 General information on unit adjustment and calibration
Examples of image acquisition with precision and coarse adjustment
Sensor adjustment can usually be performed directly via precision
adjustment. Only in exceptional cases, e.g. if one or several needles are
completely outside of the image section in an image acquired with the
"Fine adjustment"
presetting (A), is it necessary to perform coarse
adjustment prior to precision adjustment (B).
On the exposure with coarse adjustment (B), the center pin is just barely
visible on the left margin of the image section in the middle. Even in this
extreme case, an adjustment would still be possible.
7 Unit adjustment and calibration
171