Sirona ORTHOPHOS SL 2D Service Manual page 329

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
The sub menus of the
structure tree.
8. Select the desired service routine in the structure tree, e.g. S008.3.
The service routine, e.g. S008.3 is displayed.
NOTE: In terms of entry or setting options, the service routines of
the Web service correspond to the service routines that can be
called using the control panel (see chapter "Perform service
routines via the control panel [ → 240]").
9. Enter the desired parameters in the service menu and confirm the
entries by clicking the
The settings are accepted.
10. Follow the same approach for the next service routine.
11. Once you have made all of the necessary settings, exit the Web
service again.
To do this, click on the
"Service Exit"
and then on the
IMPORTANT
Until the connection to the web service is removed using the
Exit"
button, the operation of the unit with the user interface remains
blocked.
8.2 Running service routines via the web service
"Service"
menu are displayed in the
"Submit Changes"
button.
"Service DONE"
button in the structure tree
button in the following menu.
8 Service routines
"Service