Checking the CAN bus with the diagnostic function of
board DX1
Board DX1 features a diagnostic function for diagnosing malfunctions of
the CAN bus via LEDs V700 and V701 (see wiring diagrams). The
following table indicates the operating status of the CAN bus and the
recommended error correction measures:
CAN bus operation
CAN bus OK
CAN error, no communication with
board DX7, i.e. no display of error
messages
CAN error, no physical communication
with CAN bus possible; there is
probably a short circuit in the CAN
cable or on the board of a module.
CAN error, CAN bus TTP (time trigger
protocol) disturbed by defective,
constantly transmitting board (bus-
heavy).
System did not power up (DX11)
6.2.2

Jumper positions in the CAN bus

The jumpers are on board DX1 of the following sockets (see also wiring
diagrams (REF 64 95 233):
● X302 (DX7, control panel)
● X306 (DX61, diaphragm)
● X309 (DX91, cephalometer)
● X500 (DCS sensor)
● X503 (CEPH sensor)
If a cable is plugged into the socket, the corresponding jumpers must be
set to the outer position.
If no cable is plugged in, the the jumpers must be set to the inner position.
If a jumper is set to the inner position without a cable plugged in, the CAN
bus is interrupted at this location. Modules located behind this location
can no longer be connected to the CAN bus and, therefore, do not
function.
6.2 Checking the CAN bus
Error correction
Not required
● Check cabling.
● Check CAN jumper (jumper
positions in CAN bus) [ → 145]
● Disconnect CAN cables one after the
other (set jumpers to inner position!)
until the CAN bus functions again
(V700 and V701 flash slowly).
Replace faulty module.
● Disconnect CAN cables one after the
other (set jumpers to inner position!)
until the CAN bus functions again
(V700 and V701 flash slowly).
Replace faulty module.
Switch unit off and on again and wait until
end of power-on time.
6 Troubleshooting
145