Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
Pan - quality test exposure (2D)
10.11.2.3
You can use the
exposure with full rotation. This function enables you to simulate a quality
test, for example, in a similar way to an acceptance/constancy test.
1. Insert the needle phantom (A) into the bite block holder.
2. Plug the contrast element (B) into the slot on the needle phantom
provided for that purpose.
NOTICE
The aluminum plate of the contrast element must be facing away from
the column of the unit.
Do not attach any additional aluminum plate to the unit.
3. Press the R key.
The unit moves to its starting position.
4. Call the
5. In the structure tree, under
- quality control exposure"
The menu for creating the
appears in the action window.
"Pan - quality control exposure"
"Control exposures"
menu [ → 428].
"2D control exposures"
element.
"Pan - quality control exposure"
10 Maintenance
10.11 Test exposures/Test images
to create a panoramic
"Pan
, select the
431