Checking the tube data

10.10.1

Checking the tube voltage

Preparing the measurement
1. Attach the Mult-O-Meter sensor in the middle of the sensor (on 2D
sensor side).
2. Set the main switch (A) to I (see also Operating Instructions).
3. Wait for approx. 1 minute.
4. Touch the R key on the touchscreen.
The unit moves to its starting position.
Performing measurements
1. Select the P1 program (see Operating Instructions).
2. Set the kV/mA level 63kV/8 mA (see Operating Instructions).
3. Establish receptivity (see SIDEXIS 4 Operator´s Manual).
4. CAUTION! Activating the release button triggers X-rays.
Start the exposure by pressing the release button. Hold down the
release button until image acquisition is completed and the acoustic
signal that indicates the end of the exposure can be heard.
Analyzing measurements
➢ Read the voltage values on the display of the Mult-O-Meter.
IMPORTANT
The measured tube voltage must correspond with the tube voltage set
of 63kV. The permissible tolerance is ± 10 %.
If the measured values
range
If the measured values are within the permissible tolerance
range, finalize the measurement.
Concluding the measurement
➢ Switch the unit on via switch (A) (see also Operating Instructions).
are not within the permissible tolerance
tube assembly
, replace the