7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
218
Adjustment value S1 is calculated automatically from the
exposure and entered in the text box of the
main beam direction"
A
Adjustment OK
B
Adjustment not OK
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
"Adjustment of Ceph
menu.
● A horizontal bar must be visible in the
center of the image. If this bar is visible,
the exposure is OK, and the two beams
on the image are within the tolerance
range of ±10 mm.
D3632.076.01.03.02 09.2016
64 95 258 D3632