S037: Test Step 3 - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016

S037: Test step 3

8.1.21.3
Configuring boot mode
Selection field
1
* Factory setting
1. Call service routine S037.3.
Once the service routine has been selected, the current boot
mode of the unit is displayed in selection field 1.
2. Use arrow keys (A) to select the required boot mode "automatic
address assignment" (DYNAMIC) or "fixed address" (STATIC) in
selection field 1 (see table).
The Memory key (B) lights up.
3. Save the setting.
4. Exit the service routine.
8.1 Perform service routines via the control panel
Parameters
Meaning
DYNAMIC
Automatic address assignment
(DHCP/AutoIP)
STATIC
Fixed address*
8 Service routines
323