Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Error code
Description
E3 13 40
Release signal applied during
power-on.
Error code
Description
E6 13 42
The hardware signal for radiation
release is applied on board DX1
during unit operation although no
actuated X-ray release button is
being reported via the CAN bus.
Error code
Description
E5 13 43
The door was opened during the
exposure.
Error code
Description
E5 13 50
Forehead support; has not left home
position despite movement pulses.
Error code
Description
E5 13 51
Forehead support; has not left home
position and no motor movement.
64 95 258 D3632
D3632.076.01.03.02 09.2016
Actions required
E3 13 40
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure and observe causal
error messages.
If the error occurs repeatedly ...
● Check the X-ray signal path.
Actions required
E6 13 42
● Check the X-ray signal path.
Actions required
E5 13 43
● Check the X-ray signal path.
Actions required
E5 13 50
● Check light barriers V5_1 to V5_4, replace
headrest and light barrier if necessary.
● Check cable L18, replace if necessary.
● Check board DX5, replace if necessary.
Actions required
E5 13 51
● Check motor M5_1, replace headrest if
necessary.
● Check light barriers V5_1 to V5_4, replace
headrest and light barrier if necessary.
● Check cable L18 and replace if necessary.
● Check board DX5, replace if necessary.
5 Messages
5.4 List of error messages
see
S. [ → 156]
see
S. [ → 283]
see
S. [ → 283]
see
S. [ → 151],
S. [ → 347],
S. [ → 393],
S. [ → 153],
S. [ → 400],
S. [ → 146],
S. [ → 395]
see
S. [ → 150],
S. [ → 393],
S. [ → 151],
S. [ → 393],
S. [ → 153],
S. [ → 400],
S. [ → 146],
S. [ → 395]
105