If threads B should tightened:
6.
Slightly loosen the screws (B) of the laser area display.
7. Adjust the laser area display (A) vertically in such a way that the
distance between the displayed laser line and cover joint is 5 mm.
Following the adjustment, two horizontal laser lines from the laser
B
area display and a vertical laser line from the MS laser light
localizer should be displayed on the sensor unit cover.
IMPORTANT
A third line from the horizontal laser line of the FH laser module must not
be displayed. If a third horizontal laser line (FH) is displayed on the
B
sensor cover instead of the vertical laser line (MS), then the plug-in
positions on board DX1 have been swapped. In this case, swap the
plug-in positions X607 and X611 on board DX1.
8. Re-tighten the screws (B) of the laser area display.
9. Switch the laser light localizers off again.
10. Switch the X-ray unit off again.
9 Repair
9.7 Laser light localizers