Sirona ORTHOPHOS SL 2D Service Manual page 389

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Q
64 95 258 D3632
D3632.076.01.03.02 09.2016
14. Place the two new heat conducting foils (Q) (included in the scope of
supply for board DX83) on board DX83.
Q
15. Reattach the heatsink with the four screws (J) and plain washers on
the DCS sensor.
NOTE: Make sure that the HF shield on the side is positioned
correctly. The combed parts of the shield (S) must not become
trapped.
16. Refit the panel (I) with the two screws (H) on the DCS sensor (see
removal).
17. Plug the cable of the fan package back into connector X1000 on
board DX83.
18. Fit the DCS sensor into the holes (A) of the rotation unit and secure
the sensor back on the rotation unit with the 4 screws (B) (see
removal).
19. Fit the flat panel detector (or counterbalance) into the guide of the
rotation unit and secure the it back on the rotation unit with the 3
screws (G) (see removal).
20. Reconnect the cables of the sensor unit. Follow the instructions on
connecting the sensor unit in chapter "Connecting a sensor unit
[ → 380]".
21. Reattach the covers of the sensor unit (D and E) with the 2 screws (F)
(see removal).
22. Restore the previously installed sensor data using the menu
/ Restore sensor data"
the unit (refer to chapter "Restoring sensor data [ → 74]").
23. Perform a complete unit adjustment and calibration.
"Restore sensor data"
9 Repair
9.12 Sensor unit
"Backup
on board DX83 of
389