7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
210
7.2.1.6.3
Mechanical adjustment: Ceph secondary diaphragm
1. Loosen screw (A) and remove the cover of the secondary diaphragm
by pulling it downward.
2. Loosen screws (B) slightly (approx. 2-3 turns).