Service Manual ORTHOPHOS SL
DX1
64 95 258 D3632
D3632.076.01.03.02 09.2016

Replacing board DX11

9.15.2.2
The cover plate of board DX11 must be removed [ → 396].
1. CAUTION! Touch a ground point to discharge static electricity before
touching any boards.
Pull the defective DX11 board using the removal tool (D) to remove it
from the DX1 board.
IMPORTANT: The removal tool (D) is included in the delivery scope
of the DX11 board.
2. NOTICE! Ensure that the connector strips of boards DX1 and DX11
are aligned precisely above one another and are not offset, before
D
pressing the boards together firmly.
Attach the new DX11 board on to the DX1.
3. Check that the DX11 board is correctly attached to the DX1 board.
DX11
There must be no visible gap between the connector strips.
4. Reattach the cover plate.
After replacing board DX11, be sure to observe the instructions in the
section "Measures following replacement of boards [ → 398]".

Replacing board DX32

9.15.2.3
The removal of the board DX32 is described in the chapter "Removing
board DX32 [ → 335]". Install the board by following the same procedure
in reverse order.
After replacing board DX32, be sure to observe the instructions in the
section "Measures following replacement of boards [ → 398]".