S017: Test step 6
8.1.14.4
Activate / deactivate the remote control display
Selection field
1
* factory settings
1. Call service routine S017.6.
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
2. Use the arrow keys (A) to select the code for the required setting in
selection field 1 (see table).
Once the required setting has been selected, the Memory key (B)
lights up.
3. Save the setting.
4. Exit the service routine.
8.1 Perform service routines via the control panel
Code
Function
00
Display inactive*
01
Display active
8 Service routines