5 Messages
5.4 List of error messages
Error code
Description
E1 11 20
The calibration data on the unit is
invalid or does not match the serial
numbers of the modules.
Error code
Description
E6 11 21
The calibration data could not be
determined on the unit.
Communication with the frame
grabber is interrupted.
Error code
Description
E7 11 30
Occurs after service tasks in which
the device configuration, calibration
or adjustment were adapted.
Service functions continue to be
available.
Error code
Description
E1 11 88
The unit is in demo mode.
Occurs when the unit is switched on.
Error code
Description
E6 12 01
CAN controller initialization error on
DX1.
Error code
Description
E6 12 02
CAN malfunction (cannot be
assigned to module).
96
Actions required
E1 11 20
● Perform unit adjustment and calibration.
● Check the serial numbers of the flat panel
and board DX6, DX11, DX61 and frame
grabber.
Actions required
E6 11 21
Actions required
E7 11 30
For patient exposures...
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure
Actions required
E1 11 88
If user mode is expressly required ...
● Turn off the demo mode.
CAUTION! With the demo mode disabled,
radiation is triggered by pressing the
trigger button!
5.4.5

Location 12: CAN bus

Actions required
E6 12 01
● Check the CAN bus.
Actions required
E6 12 02
● Check the CAN bus.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
see
S. [ → 160]
see
see
see
S. [ → 78]
see
S. [ → 141]
see
S. [ → 141]
64 95 258 D3632
D3632.076.01.03.02 09.2016

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos sl 3dOrthophos sl 3d ceph

Table of Contents