8 Service routines
8.1 Perform service routines via the control panel
262

S005: Test step 5

8.1.6.3
Temperature sensor test, single tank
1. Call service routine S005.5 [ → 249].
After the service routine has been selected, selection field 1
displays the single tank temperature in °C. The display is updated
once per second.
2. Exit the service routine [ → 252].
8.1.6.4

S005: Test step 6

Switch off cool-down interval of the single tank
NOTICE
Risk of X-ray tube assembly failure
Running this service routine is irreversible!
The device is operated beyond its specifications afterwards. This may
result in an X-ray tube assembly failure. As a result, the system warranty
expires. SIRONA is able to prove this system setting after a system
failure.
IMPORTANT
Upon explicit customer request, the monitored pulse/pause ratio can be
disabled in exceptional cases. This can only be done once by running
this service routine. Re-enabling the pulse/pause monitoring function is
not possible, though. Such a one-time disablement can be displayed on
the info screen and in the detail query by entering "PulsePause-
Disable".
Selection field
1
* factory settings
1. Call service routine S005.6 [ → 249].
The status of the pulse/pause treatment is displayed in selection
field 1 (see table).
Service Manual ORTHOPHOS SL
Code
Function
00
Pulse/pause monitoring inactive
01
Pulse/pause monitoring active*
Sirona Dental Systems GmbH
64 95 258 D3632
D3632.076.01.03.02 09.2016