Service Routines; Perform Service Routines Via The Control Panel; Overview Of Service Routines; List Of All Service Routines Available For Selection - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

8 Service routines
8.1 Perform service routines via the control panel
Service routine
S001
S002
S002.1
S002.3
S002.4
S005
S005.2
S005.4
S005.5
S005.6
S005.7
S007
S007.1
S007.2
S008
S008.2*
S008.3*
240
8

Service routines

IMPORTANT
sirona-internal SRs
Service routines that are displayed on the unit control panel but are not
described in the technical documentation are purely for Sirona-internal
purposes and may not be executed under any circumstances.
Executing these service routines may result in a malfunction in the unit
or parts of the unit.
8.1

Perform service routines via the control panel

NOTE: This section describes how to perform the service routines via the
Note on the web service
service menu of the control panel.
The service routines S008.3, S017.2 - S017.18 and S037.4 can also be
run via the Sirona web service (see section Running service routines via
the web service [ → 327]).
8.1.1

Overview of service routines

List of all service routines available for selection

8.1.1.1
Function
Radiation without rotary movement, fixed maximum radiation time S. [ → 253]
Radiation without rotary movement, selectable kV/mA level and
maximum radiation time
Radiation without rotary movement; selectable kV/mA level and
maximum radiation time; primary diaphragm in last position
selected
Radiation without rotary movement; selectable kV/mA level and
maximum radiation time (max. 14s); primary diaphragm fully
opened with respect to PAN position
Radiation without rotation; maximum radiation time selectable; the
primary diaphragm is set to a widened opening symmetrical to the
PAN setting so that the step filter still remains covered
General X-ray tube assembly service
Automatic adjustment of preheating
Tube assembly fan test
Temperature sensor test, single tank
Switch off cool-down interval of the single tank
Configuring continuous operation of the tube assembly fan
Fault buffer
View error memory
Clear error memory
Update service
Overview of module software versions
Confirming the unit serial number
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
see
S. [ → 255]
S. [ → 255]
S. [ → 256]
S. [ → 258]
S. [ → 260]
S. [ → 260]
S. [ → 261]
S. [ → 262]
S. [ → 262]
S. [ → 264]
S. [ → 265]
S. [ → 265]
S. [ → 266]
S. [ → 267]
S. [ → 267]
S. [ → 267]
64 95 258 D3632