What Has To Be Done After Replacing The X-Ray Tube Assembly - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
9.11.4
What has to be done after replacing the X-ray tube
assembly?
IMPORTANT
The tube assembly may be replaced only as a complete module.
IMPORTANT
Note the dependency of the tube assembly on device serial numbers in
the spare parts list.
IMPORTANT
Use the most current software version available on the ORTHOPHOS
SL Firmware DVD.
1. Switch the unit on.
Error message E6 15 04 is occasionally displayed at this point.
This error message can be ignored at this point. Acknowledge
the error message with the R key, if necessary.
Error message E61505 (undefined unit serial number) is
displayed.
2. Acknowledge the error message using the R key on the control panel.
3. Call the service menu.
4. Call up service routine S008.3.
5. Check the serial number
IMPORTANT: The unit serial number is located on the nameplate of
the unit.
NOTICE! If the serial number is incorrect, cancel the update and
contact the Sirona Customer Service Center!
6. Confirm the serial number using service routine S008.3.
7. Perform a software update to the current software version.
All modules are updated in accordance with the configuration.
8. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
9. Perform a complete unit adjustment and calibration.
10. Delete the X-ray component in the SIDEXIS 4 configuration menu
"ORTHOPHOS SL"
(refer to chapter "Setting up the X-ray component [ → 49]").
Only for Germany, Austria, Switzerland and France:
11.
Perform an acceptance check without consulting an expert.
The process is completed.
"Device Administration"
/
9 Repair
9.11 X-ray tube unit
and add a new one
377