9 Repair
9.15 Boards
398
9.15.3

Measures following replacement of boards

IMPORTANT
After replacing boards or modules containing boards, check to make
sure that the software version of the module corresponds to the current
software version of the unit.
The software versions of the modules can be queried by running service
routine S008.2 or using the
can also check the info screen in advance to determine whether the
current software constellation is permissible. If this is not the case, the
version number of the entire software is marked by an asterisk (e.g.
V04.03.01*).
In case of software incompatibilities, carry out a software update or
downgrade.
Always carry out the measures described below exactly in the order
specified and do not perform any other actions in between.
9.15.3.1

After changing the DX11 board

1. Switch the unit on.
NOTICE! Do not acknowledge any error messages at this time.
2. Call [ → 76] the info screen.
A "Factory Default" DX11 must be displayed on the info screen.
3. Call the service menu [ → 248].
4. Call service routine S017 [ → 249] and perform [ → 278] service
routine S017.2:
5. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
NO key and the error message E1 11 20 are displayed.
"Automatic"
6. Use the
update to the current unit software version.
The current update file is located on the
DVD"
on the Sirona dealer page on the Internet.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
"Device details"
in SIDEXIS 4[ → 59]. You
update mode to perform an automatic software
"ORTHOPHOS SL Firmware
D3632.076.01.03.02 09.2016
64 95 258 D3632