9 Repair
9.10 Diaphragm unit
370
9.10
Diaphragm unit
9.10.1
Replacing the diaphragm unit
IMPORTANT
Only the diaphragm unit of ORTHOPHOS SL may be installed! Note the
dependency of the diaphragm unit on device serial numbers in the spare
parts list.
IMPORTANT
The diaphragm unit may be replaced only as a complete module.
1. Remove the "Rear tube assembly" cover [ → 37].
2. Pull cables L11 and L12 off connectors X501 (L11) and X101 (L12)
on board DX61.
3. Pull off the cables L21 (2x) and L20 from the connectors X303, X304
(L21) and X203 (L20) on board DX61.
A
A
4. NOTICE! Hold on to the diaphragm while you remove it.
Loosen the screws (A) on the old diaphragm unit and remove it from
the X-ray tube assembly.
5. NOTICE! Hold on to the diaphragm while you attach it.
Mount the new diaphragm unit with the screws (A) on the X-ray tube
assembly.
6. Plug cables L20 and L21 back into connectors X203 (L20), X302
(L21) and X303 (L21) on board DX61.
7. Plug cables L11 and L12 back into connectors X501 (L11) and X101
(L12) on board DX61.
8. Reattach the "Rear tube assembly" cover [ → 37].
A
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016