4 General operating procedures
4.5 Updating the unit firmware
Starting the software update
68
4.5
Updating the unit firmware
NOTICE
Firmware update
Unit inoperability!
Before starting the firmware update, make sure that no unit movements
are active. Otherwise the system may become inoperable in rare cases.
The sensor unit must be installed as part of the update. Exposure
readiness must be deselected in SIDEXIS 4 and the unit must not
already be in service mode.
IMPORTANT
Read the information provided on the firmware CD supplied with the unit
and on the SIRONA dealer page on the Internet very carefully. These
sources always contain the latest information on software updates.
✔
SIDEXIS 4 is installed.
✔
The workstation software for the corresponding X-ray component or
sensor is installed on the workstation.
1. Start SIDEXIS 4.
2. In SIDEXIS 4, call the
configuration menu.
3. Select the desired X-ray component in the list marked
administration"
4. Then click on the
Service Manual ORTHOPHOS SL
"Orthophos SL"
"Device Administration"
/
.
"Software update"
button.
Sirona Dental Systems GmbH
"Device
64 95 258 D3632
D3632.076.01.03.02 09.2016