Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Error code
Description
E5 13 73
Malfunction of height adjustment
during operation.
Error code
Description
E5 13 83
Error while generating pulse for
sensor.
Error code
Description
E6 13 87
Error when activating pulse
generation.
64 95 258 D3632
D3632.076.01.03.02 09.2016
Actions required
E5 13 73
● Acknowledge error.
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch on the unit.
4. Repeat procedure and check function.
● Please contact the Sirona Customer
Service Center (CSC) to find out whether
a bug fix by means of a software update is
possible and perform such an update if
necessary.
If the error occurs repeatedly ...
● Replace board DX1.
Actions required
E5 13 83
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch on the unit.
4. Repeat procedure and check function.
● Please contact the Sirona Customer
Service Center (CSC) to find out whether
a bug fix by means of a software update is
possible and perform such an update if
necessary.
If the error occurs repeatedly ...
● Replace board DX1.
Actions required
E6 13 87
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch on the unit.
4. Repeat procedure and check function.
● Please contact the Sirona Customer
Service Center (CSC) to find out whether
a bug fix by means of a software update is
possible and perform such an update if
necessary.
5 Messages
5.4 List of error messages
see
S. [ → 68]
S. [ → 396]
see
S. [ → 68]
S. [ → 396]
see
S. [ → 68]
109