What has to be done after replacing the DCS sensor?
9.12.2.3
IMPORTANT
Use the most current software version available on the
SL Firmware DVD"
1. Reattach the covers of the sensor unit in reverse order of removal.
2. Switch the unit on.
3. Check the rotation unit for a smooth rotation by switching between the
operating modes
Only if there is a flat panel detector:
4.
Restore the previously installed sensor data using the menu
/ Restore sensor data"
the unit (refer to chapter "Restoring sensor data [ → 74]").
5. Perform a firmware update to the current unit software version.
6. Perform a complete unit calibration (see the chapter titled
"Adjustment and calibration via the calibration menu [ → 181]").
7. Delete the X-ray component in the SIDEXIS 4 configuration menu
"ORTHOPHOS SL"
(refer to chapter "Setting up the X-ray component [ → 49]").
8. Create a backup copy of the sensor data (see chapter "Backing up /
restoring / installing sensor data [ → 71]").
Only for Germany, Austria, Switzerland and France:
9.
Perform an acceptance check without consulting an expert.
.
Pan
3D
and
via the control panel.
"Restore sensor data"
⇒
"Device Administration"
/
9 Repair
9.12 Sensor unit
"ORTHOPHOS
"Backup
on board DX83 of
and add a new one