Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
5. Secure the contrast element using a hygienic cap.
6. Attach the ceph test phantom to the clip provided for that purpose.
7. Push the ceph test phantom with the clip from below on to the
secondary diaphragm of the ceph arm.
IMPORTANT
The clip must engage into the opening on the secondary diaphragm
provided for that purpose.
8. Press the R key.
The unit moves to its starting position.
10 Maintenance
10.11 Test exposures/Test images
439