10 Maintenance
10.10 Checking the tube data
422
10.10.2

Checking the tube current

NOTICE
Damage to the measuring unit
The ring assembly and the tube assembly move during the
measurement.
Make sure that the measuring wires are sufficiently long to allow for the
ring movement and that the measuring unit is in a secure position so that
it will not fall down.
Preparing the measurement
1. Switch off the unit (see Operating Instructions).
DANGER! After switching off the unit, wait at least 4 minutes (LED
V500 on the DX6 must no longer be on) before removing the cover
on the tube assembly.
2. Loosen the screw (D) and remove the lid of the tube assembly cover
(E).
D
E
3. Loosen the 4 screws (F) and remove the cover plate (G).
F
G
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016