7.1 General information on unit adjustment and calibration
Password protection
The calibration menu is password-protected. For the password, enter the
first four digits of the current system date (PC) in reverse order. Example:
On 25.03.2016, the service password is 3052.
Service mode
When you open the calibration menu, the unit switches from user mode
to the PC service mode blocked by the PC. In PC service mode, the
control options that are available on the control panel are determined by
SIDEXIS 4 and the service routine currently selected. General control of
the unit by means of the control panel (as in the user mode) is not
possible in this mode.
Service mode is displayed on the Easypad via the PC service image.
7 Unit adjustment and calibration