Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS XG
5
Program group panoramic images
5.1
P1 standard panoramic image, P1 A artifact reduced,
P1 C with a constant magnification factor of 1.25
PAN
CEPH
TS
1260
P1
SID = 19,6"
SID = 19,6"
10
9,0 s
9,0 s
6 x 12"
6 x 12"
Quick
Ready for exposure
Ready for exposure
Active sensor area CCD PAN
6.5 x 138.4 mm,
for upper or lower jaw:
6.5 x 70.0 mm
59 87 594 D 3352
D 3352.201.01.18.02
Plus
DS/Ceph
5.1 P1 standard panoramic image, P1 A artifact reduced, P1 C with a constant
6
8
3
64kV
8mA
?
5 Program group panoramic images
To preselect the program, use the arrow keys (3): + (to
count forward) or – (to count backward), depending on
the starting position; for program sequence see section
3.2.
To preselect the exposure type, touch the program dis-
play (6). A continuous loop will then display all programs
one after another, P1, P1 A, P1 C, P1...
In the "Half-view and individual quadrant selection" sub-
menu at the top of column (8) you can preselect whether
you want to generate a right or left half-view, or a maxil-
lary or mandibular view (or an individual quadrant view
in the full version) (see section 3.2 General touchscreen
functions).
P1 Normal view
•
Yellow bite block or contact segment
or
chin rest with bite block rod and bite block or bar.
•
Head inclination using the FH.
The line displayed on the touchscreen head icon in-
dicates the reflection line of the light localizer.
•
Optional presetting for malocclusion
•
Optional presetting for jaw shape
•
Automatic selection of the slice width in accor-
dance with temple support settings for different den-
tal arches. Radiation time depends on set temple
support width.
P1 A normal view, artifact-reduced
•
Settings as for P1
•
To prevent metal artifacts in the condylar and molar
region and to reduce shadowing caused by the op-
posite jaw.
P 1C Normal view with a constant
magnification factor of 1.25
e.g. for implantology
•
Settings as for P1
31