Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Error code
Description
E4 13 01
Actuator 1 has not reached pan
home position.
Error code
Description
E4 13 02
Actuator 1 has not left pan home
position.
Error code
Description
E5 13 03
Malfunction of actuator 1 during
operation.
64 95 258 D3632
D3632.076.01.03.02 09.2016
5.4.6
Location 13: Stand/Peripherals
Actions required
E4 13 01
● Check the actuator mechanisms manually
for smooth and easy running.
In case of stiffness ...
● Manually reset the rotary ring to the pan
home position and run a test cycle via P1.
If the error occurs again ...
● Check light barrier V1_1 (X802) with
service routine S015.5, replace light
barrier if necessary.
● Run service routine S015.5 (free travel) to
check electronic activation and jerk-free
running, replace actuators and/or spindles
and board DX1 if necessary.
Actions required
E4 13 02
● Check the actuator mechanisms manually
for smooth and easy running.
In case of stiffness ...
● Manually reset the rotary ring to the pan
home position and run a test cycle via P1.
If the error occurs again ...
● Check light barrier V1_1 (X802) with
service routine S015.5, replace light
barrier if necessary.
● Run service routine S015.5 (free travel) to
check electronic activation and jerk-free
running, replace actuators and/or spindles
and board DX1 if necessary.
Actions required
E5 13 03
● Check the actuator mechanisms manually
for smooth and easy running.
In case of stiffness ...
● Manually reset the rotary ring to the pan
home position and run a test cycle via P1.
If the error occurs again ...
● Check light barrier V1_1 (X802) with
service routine S015.5, replace light
barrier if necessary.
● Run service routine S015.5 (free travel) to
check electronic activation and jerk-free
running, replace actuators and/or spindles
and board DX1 if necessary.
5 Messages
5.4 List of error messages
see
S. [ → 276],
S. [ → 330]
see
S. [ → 276],
S. [ → 330]
see
S. [ → 276],
S. [ → 330]