Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
2. Once you have taken the exposure, measure distances S1, S2 and
S3 with the toolbar measuring ruler [ → 166].
3. Overwrite the adjustment values for S1, S2 and S3 in the text boxes
"Adjustment of Ceph secondary diaphragm"
of the
measured values. For information about the shifting direction (input
of +/- sign in the menu), see the section entitled "Shifting direction of
the exposed image area [ → 174]".
NOTICE! Use points as decimal separators!
4. Continue with the adjustment procedure as described in the chapter
entitled "Automatic adjustment: Ceph secondary diaphragm
(standard) [ → 205]"
7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
menu with the