8. Establish [ → 175] receptivity.
9. Take an exposure (85 kV/22 mA s) [ → 175].
A message is displayed in the action window to indicate whether
adjustment has been successful and the status indicator in front
"5x5"
"Premolar left"
of the
/
responds accordingly:
Calibration not OK = status indicator stays red
Calibration OK = status indicator turns green
NOTE: The calibration values are automatically saved.
10. Repeat the procedure starting at Step 8 until calibration is successful.
11. Repeat calibration for all centers of rotation of volume
all other volumes and their centers of rotation.
12. Remove the 3D calibration block from the pan bite block holder on the
unit.
7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
element in the structure tree
"5x5"
and for