7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
232

Geometry calibration

7.2.2.5
1. Insert the 3D calibration block into the bite block holder on the unit.
[ → 179]
2. Align the 3D calibration block with the spirit level by adjusting the
screw (A).
* For greater clarity, the unit in the left image is displayed without
*
sensor unit.
3. In the structure tree, under
click on the
"Molar left"
The
4. Establish [ → 175] receptivity.
5. Create an exposure (85 kV / 22 mA s) [ → 175].
A message is displayed in the action window to indicate whether
adjustment has been successful and the status indicator in front
"5x5"
of the
A
accordingly:
Calibration not OK = status indicator stays red
Calibration OK = status indicator turns green
NOTE: The calibration values are automatically saved.
6. Repeat the procedure starting at Step 4 until calibration is successful.
7. Click on the
"5x5"
volume
"3D"
"Geometry calibration"
/
"Molar left"
/ element (S010.25 - S010.39).
element is selected.
"Molar left"
/
element in the structure tree responds
"Next"
button or select the next center of rotation of
.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
"5x5"
/
64 95 258 D3632
D3632.076.01.03.02 09.2016
,