4.6.4

Installing sensor data

Sensor data (factory calibration of the sensor) of a new sensor can be
installed on the DX83 board using the
IMPORTANT
After the sensor data has been installed, the unit adjustment and
calibration must be performed [ → 160] in full.
If no complete unit adjustment and calibration has been performed, no
patient images are possible.
NOTE: As the sensor data is stored on the DX83 board, this is necessary,
for example, after changing the flat panel detector. The corresponding Zip
file is included in the scope of supply of the new flat panel detector.
"Backup / Restore sensor data"
The
1. In the structure tree, click on the
"Install new sensor data"
The
2. Click on the
contains the sensor data for the new flat panel detector in the
directory.
3. Click on the
The sensor data stored in the Zip file is installed on the DX83
board.
4. Perform a complete unit adjustment and calibration [ → 160].
4.6 Backing up / restoring / installing sensor data
"Install new sensor data"
menu is opened.
"Install new sensor data"
menu is displayed.
"Select File"
button and select the desired Zip file that
"Install"
button.